Henan Forever Medical CO., LTD. Company Logo Henan Forever Medical CO., LTD.
5 1

Hospital Laboratory Clinical Chemistry and Biochemistry Analyzer

Contact Now* Send an Inquiry to this supplier.
or
Start Order * Name your price
You may also be intersted in
biochemistryanalyzerbiochemistry analyzer

Contact us

Henan Forever Medical CO., LTD.
[China]

Address
Wanda Center, Jinshui District, Zhengzhou, Henan, China Zhengzhou Henan
Phone
86-0371-61315851
Contact name
Judy

Quick Information

  • Brand Name: forered
  • Place of Origin: China
  • Model Number : YJ-SL180
  • Net Weight : 60KGS
  • Classificatio : Biochemical Apparatus
  • Humidity : More Than 70%
  • Power Supply : AC 220V±22V 50Hz&p
  • Frequency : Maximum 600va

Description

High End Full Automatic biochemistry analyzer Specification:  Throughput constant 180 T/h or 300T/h,best for middle or small size hospital
  Closed integrated photovoltaic systems, no mechanical parts,maintenance free
  Minimum reaction volume is 150ul,cost saving
  Discrete,random access,automatic
  Random choice of method, the real discrete automatic chemistry analyzer
  This automatic chemistry analyzers have gained a lot of national patents during the researching.
  Analyze Unit1.  Throughput: constant 180 T/h, 81 reaction cuvette positions,multi-channel testing together.
2.  Minimum reaction volume is 150μl, cost saving.
3. Determined by the sample sequence, inserted at any time emergency.
4.  Heating method: Direct thermal solid thermostat closed.Optical Unit1.  Closed integrated photovoltaic systems, no mechanical parts,maintenance free.
2. Light Source:Halogen-tungsten lamp;
3. Lifetime≥2000 h.
4. 8 Wavelength,support single or dual wavelength.Reagent Unit1.  60 reagent positions.
2.  Automatically detect the amount of reagent remaining, displayed in real time the remaining amount of reagent.
3.  15ml, 20ml, 30ml, 50ml or 70ml reagent bottle. Dead volume <1.0 ml.
4.  Support for single and double reagent, reagent completely open.
Internal barcode reader (option).Sample Unit1. 60 sample positions.
2.  Testing according setting,random access.
3.  Real time display of testing status.
4.  Internal barcode reader (option).Reagent/Sample handing Unit1. Intelligent liquid detection & collision protection (Red light: collision; Blue light: detection of the liquid).
2.  Reagent volume: 2 ~ 500μl,step by 1μl .
3.  Probe cleaning:automatically washing for both inside and outside,carry over0.1%.
4. Automatically dilute and retest.Calibration and quality control Unit1. Linear curve type:Single-point linear,two-point linear and multi-point linear.
2. Nonlinear curve type:Logit-4p,Logit-5p,polynomial,factor.
3.  Automatically use best calibration curve and standard.
4.  Quality control chart:L-J,Westgard TwinPlot,Cumulative.Operation Unit 1.  Operation system: Windows XP/Office 2000,interface is intuitive and operate it easily, arbitrary operating authority;No restrictions on the amount of time and storage, permanent preservation of various types of data, automatic storage and backup.
2. Can print the contents of any custom;Offers a variety of print formats.
3.  Help and operation guide on line in English.
4.  No restrictions on setting test parameters.
5. Unique password for user management,rational allocation of user rights.
6.  Real time display of testing status,the remaining amount of reagent,testing status ,reaction curve and the test progress.
7. Real time display of reaction plate temperature.
RS-232 standard interface,support for the hospital network, print reports remotely.
 Others1. Temperature:10-30 degree
2. Humidity:less than 70%.
3. Try to avoid strong light.
4.  Floor level up:less than 1/200.
5. Barometric:86.0kPa-106.0kPa.
6.  Power Supply:AC 220V±22V 50Hz±1 Hz.
7.  Water consumption:less than 6L/h.
8. Frequency:550 VA.
9.  Dimension:782mm×581mm×545mm
10. Weight:60kg
 Test projects1. ALT,AST,TP,ALB,TBIL,DBIL,ALP,r-GT,CHE,TBA,5-NT
2.  UREA,CR,UA,TCO2
3. CHOL,TG,HDL,LDL,ApoA,ApoB,LPa
4. AMY,LIP
5. GLU,GSP
6. CK,CK-MB,a-HBDH,LDH,MB
7. CRP,RF,ASO

Model YJ-180
Analyze Unit Throughput Constant 180T/h (non-ISE)
Theory Absorption spectrometry, transmission turbidimetry
Methodology End-point \Fixed-time\Kinetic\Ion selective electrode
Test according set up, random access Reaction plate Contain 81 reaction cuvettes
Maximum reaction time 15 minutes
Minimum reaction volume 150 ul
Material of cuvettes Special UV-transparent optical glass
Cuvettes use Automatically cuvettes washing
Optical Unit Light Source Halogen-tungsten lamp
Photometer Reversed optics
Wavelength 340nm-900nm wavelength option
Absorption 0-4 Abs
Reagent Unit Reagent plate 60 reagent positions ,support for single and double reagent, reagent completely open
Reagent storage temperature 2~10 degree
Reagent bottle 15ml, 20ml, 30ml, 50ml or 70ml ,dead volume is less than 1.0 ml
Reagent volume 20-500 ul, step by 1.0 ul
Reagent/Sample Probe Automatically checked ;Patent technology liquid level detection;Collision protection probe cleaning:Automatic washing for both inside and outside.
Sample Unit Sample plate 60 sample positions
Sample type Human (animal) serum \plasma\urine etc.
Samples cup Support one-off micro cups
Sample volume 2-100 ul, step by 0.1 ul
Sample/reagent probe Automatically checked ;Patent technology liquid level detection;Collision protection probe cleaning:Automatic washing for both inside and outside.
Temperature control Unit Heating method Direct thermal solid thermostat closed
Temperature equilibrium time <3 minutes
Reaction temperature 37 ±0.1 degree
Temperature fluctuation ±0.1 degree
Operation Unit Software WindowsXP \2000
Database SQL Server
Data Storage Automatic store & copy; No time and volume limit; Lifetime preservation
Print format Can be set up, offer several standard format
Calibration mode Linear curve type: Single-point linear, two-point linear and multi-point linear none linear; Curve type: Logistic-Log 4P,Logistic-Log5p,Spline,index automatically use best calibration curve and standard
Quality control methods : Real time, within day & between day
Quality control diagram L-J \cumulative\twin plot
Status display Real time monitoring of reagent residual volume .Running status of reagent tray, samples tray
Adjustment Dynamic and real time display of running status; Intelligent carry over settings to adjust test sequence; original reaction data record; Detail profile of alert messages       
Setting Operation permission setting, Item setting, cross pollution setting, cuvette blank setting
Others Water consumption : < 6L/h Power Supply :AC 220V±22V 50Hz±1Hz
Frequency :Maximum 600VA Dimension :782mm×581mm×545mm
Net Weight : 60kg Accessory :Software
 
Note:We provides you with 12-24 months free repair guarantee for main unit, 12 months free repair guarantee for probe and 6 months free repair guarantee for the optional parts respectively.More information for our products and service,please turn to our official website or contact us directly.
 


Certificates

  • CE : 111111111111

Price

  • Price Condition : FOB
  • Price : USD6524 / 1
  • Delivery Port : shanghai

Packaging & Delivery

  • Packaging Size : 782 x 584 x 545  Millimeters
  • Package Weight : 60  Kilograms
  • Delivery Lead Time : 15 days for ready goods
  • Minimum Order : 1
  • Supply Ability : 666set/month

Product Image

  • Hospital Laboratory Clinical Chemistry and Biochemistry Analyzer image

Send an Inquiry to this supplier

Send an Inquiry to this supplier
* From
  To Judy
Henan Forever Medical CO., LTD.
* Buying Product
- Please enter your specific buying item.
e.g. 42 inch LCD TV, leather executive chair
Category : Medical Analyzer
* Message

Use English only Max. 2000 characters. (Min. 20)

Send
EC21.com does not guarantee the validity of product information or the credentials of sellers.